Valleylab · Model Force Argon II
Improved precision and control in electrosurgery Now the benefits of precise, controlled argon-enhanced electrosurgery are available in a compact, ea Argon-enhanced coagulation offers precise energy delivery for efficient, noncontact coagulation over large surface areas. Argon-enhanced coagulation produces rapid hemostasis with a thinner, more flexible eschar and reduced rebleeding. Argon-shrouded cut facilitates cutting with less smoke and enhanced visibility. Gas flow control enhances patient safety. The low flow rate, designed to help prevent overpressurization in laparoscopic procedures, is automatically invoked at system start up. The standard flow rate can be selected for open procedures. The laparoscopic overpressure monitor helps avoid excessive peritoneal pressure in laparoscopic surgery. Audio and visual alarms warn of an overpressure condition. Flexible system design reduces space requirements and offers optimal surgical flexibility. The compact unit can be cart or ceiling mounted and is compatible with a full range of Valleylab electrosurgical generators and argon accessories. The universal mounting system is designed to accommodate several pieces of equipment efficiently and conveniently. Compatibility with the ArgonPlus™ handset allows the Force Argon™ II system to be used with this four-function handset and a variety of electrodes designed for open and laparoscopic surgery.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No cut / coag effect, "high impedance" or output-error message, output shuts down.
Root cause: Poor electrode contact or a dry field, a damaged handpiece / cable, a return-electrode (REM/NESSY) fault, or an output-stage fault.
Symptoms: REM / NESSY / plate alarm, monopolar output inhibited.
Root cause: Poorly applied or dried split dispersive pad, a pad-cable fault, or the contact-quality-monitor circuit.
Symptoms: Argon does not flow, plasma will not ignite, gas-supply or APC-probe alarm.
Root cause: Empty / low argon supply, a flow-control valve fault, or an APC-probe / connector fault.
Symptoms: Output power or the REM circuit outside tolerance on the analyser.
Root cause: Output board or REM circuit needs recalibration - anti-static handling required.
Symptoms: Output will not activate from the pedal or the handpiece.
Root cause: Footswitch or handswitch / cable fault.
Symptoms: Unresponsive touchscreen, will not power on or boot, fan / thermal alarm.
Root cause: Display, power supply, control board, or a cooling-fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibration required after replacement; anti-static handling.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reusable instrument / handpiece cables | Periodic replacement | $$ | Easy |
| Split dispersive (return) pads | Consumable | $ | Easy |
| Footswitch | Periodic replacement | $$ | Fairly easy |
| APC probe / argon flow-control valve | Periodic replacement | $$$ | Moderate |
| RF output board Recalibration required after replacement; anti-static handling. | Major service part | $$$$ | Difficult |
| Cooling fan | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.