Extracted directly from this unit's own service manual.
A 01Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A/E 02 – 04Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A/E 05Button error receptacle 1. Check the keyboard.
Check the keyboard.
A/E 06Button error receptacle 2. Check the keyboard.
Check the keyboard.
A 07 – 09Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A 0ARestart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A/E 0BInformation in test mode (adjustment).
A/E 0C – 0DRestart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A 10Underpressure at selected gas input. Check the gas supply (tank, pressure regulator). If the error persists, notify ERBE Service.
Check the gas supply (tank, pressure regulator). If the error persists, notify ERBE Service.
A 11Overpressure at selected gas input. Check the gas supply (tank, pressure regulator). If the error persists, notify ERBE Service.
Check the gas supply (tank, pressure regulator). If the error persists, notify ERBE Service.
A 12Caloric and differential pressure sensor do not agree (wrong gas). Check the gas supply (type of gas). If the error persists, notify ERBE Service.
Check the gas supply (type of gas). If the error persists, notify ERBE Service.
A 13Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A/E 21 – 23Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A 30Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A/E 38Type detection of receptacle 1 fails to agree with the stored value. Check and save the receptacle configuration in the "Version List" test program. If the error occurs again, notify ERBE Service.
Check and save the receptacle configuration in the "Version List" test program. If the error occurs again, notify ERBE Service.
A/E 39Type detection of receptacle 2 fails to agree with the stored value. Check and save the receptacle configuration in the "Version List" test program. If the error occurs again, notify ERBE Service.
Check and save the receptacle configuration in the "Version List" test program. If the error occurs again, notify ERBE Service.
A 40Flow specification not attained. Check the accessories. If the error persists, notify ERBE Service.
Check the accessories. If the error persists, notify ERBE Service.
A 41Flow specification exceeded. Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A/E 7D + 7ERestart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A/E 7FOperating system error. Information, no fault condition.
A/E 80Internal state incorrect. Information, no fault condition.
A/E 81 – 83Protocol violation CAN. Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
A/E 85Invalid resistance instrument number. Check the accessories. If the error persists, notify ERBE Service.
Check the accessories. If the error persists, notify ERBE Service.
A 86Gas underdose, e.g. hose blocked. Check the accessories. If the error persists, notify ERBE Service.
Check the accessories. If the error persists, notify ERBE Service.
A 90Low pressure at cylinder 1. Check the gas supply (tank, pressure regulator). If the error persists, notify ERBE Service.
Check the gas supply (tank, pressure regulator). If the error persists, notify ERBE Service.
A 91Low pressure at cylinder 2. Check the gas supply (tank, pressure regulator). If the error persists, notify ERBE Service.
Check the gas supply (tank, pressure regulator). If the error persists, notify ERBE Service.
A A0Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
B 02 + 03Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
B 05 (as to 1.1.x)NESSY message; NE is not correctly applied: Occurs if measurement of the NE contact resistance is outside the valid range on activation. Check the NE accessories and NE setting in setup.
Check the NE accessories and NE setting in setup.
B 0B (as from 1.2.x)NESSY message; NE is not correctly applied: Occurs if measurement of the NE contact resistance is outside the valid range on activation. Check the NE accessories and NE setting in setup.
Check the NE accessories and NE setting in setup.
B 10Please terminate activation: Occurs if activation has been automatically terminated (e.g. by AutoStop) and the activation signal remains (longer than 5 s) (e.g. footswitch). Check the accessories. If the error persists, notify ERBE Service.
Check the accessories. If the error persists, notify ERBE Service.
B 14Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
B 17Double activation: Occurs if two activation signals (e.g. both pedals of a footswitch) are present simultaneously (within 100 ms). Check the accessories. If the error persists, notify ERBE Service.
Check the accessories. If the error persists, notify ERBE Service.
B 19PowerFail: Occurs if a PowerFail signal is received but the PowerFail does not actually happen within 2 s. Check the supply input voltage.
Check the supply input voltage.
B 1CThe maximum ON time has been exceeded: Occurs if activation lasts longer than the ON time selected in Setup. Check the accessories. If the error persists, notify ERBE Service.
Check the accessories. If the error persists, notify ERBE Service.
B 1EKeyboard error: Occurs if a button pressed is recognized during initialization. Check the keyboard.
Check the keyboard.
B 1FNESSY symmetry monitoring: Occurs if an error is signaled by the Nessy symmetry monitoring during activation. Check the NE accessories.
Check the NE accessories.
B 2A – 2CRestart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
B 50 – 59Restart the unit. If the error occurs again, notify ERBE Service.
Restart the unit. If the error occurs again, notify ERBE Service.
B 5AError in the version check: Occurs if the VEM module attempts to log with an invalid version. Remove the module concerned from the unit and notify ERBE Service.
Remove the module concerned from the unit and notify ERBE Service.
B 5BError in the version check: Occurs if the IES module attempts to log with an invalid version. Remove the module concerned from the unit and notify ERBE Service.
Remove the module concerned from the unit and notify ERBE Service.
B 5CError in the version check: Occurs if the EIP module attempts to log with an invalid version. Remove the module concerned from the unit and notify ERBE Service.
Remove the module concerned from the unit and notify ERBE Service.
B 5DError in the version check: Occurs if a dual-pedal footswitch attempts to log with an invalid version. Remove the module concerned from the unit and notify ERBE Service.
Remove the module concerned from the unit and notify ERBE Service.
B 5EError in the version check: Occurs if a single-pedal footswitch attempts to log with an invalid version. Remove the module concerned from the unit and notify ERBE Service.
Remove the module concerned from the unit and notify ERBE Service.
B 5FError in the version check: Occurs if a ReMode footswitch attempts to log with an invalid version. Remove the module concerned from the unit and notify ERBE Ser
Remove the module concerned from the unit and notify ERBE Service.