Integrated Smoke Evacuation - The ESU-110 utilizes a design that integrates the electrosurgical generator and smoke evacuation system into a single compact unit. This results in an optimal design for operating areas with limited space, especially offices. It also allows simultaneous "single switch" activation of both modules by either the handswitch control pen or footswitch. Controlled Output Circuitry - Incorporates controlled output circuitry to produce the best histological tissue specimen for the pathologist. Controlled output circuitry maintains the output voltage within the ideal cutting range to produce a specimen with minimal thermal damage at the margins. It also allows the generator to be set at a preferred 'cut' power output setting, but continuously adjusts itself to meet the minimum needs for optimal cutting. This eliminates any need to adjust the output setting when changing loop sizes. Three-Stage Particle Filtration - The Generator utilizes a three-stage filtration system to evacuate and filter the smoke plume produced during electrosurgery. The filtration system includes an activated charcoal filter which absorbs odorous gases, and HEPA and ULPA filters which remove solid particles and aerosols. The 3 filters provide a minimum efficiency of 99.999% fr particles at 0.1 microns. Safety - Each system meets ANSI/AAMI HF-18 standards for patient lead isolation. This provides protection for both patient and clinician by reducing the possibility of creating an alternate current path that could result in shock or burn. In addition, the systems will sound an alarm and automatically shut off should an interruption occur in the connection to the dispersive pad.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No cut / coag effect, "high impedance" or output-error message, output shuts down.
Root cause: Poor electrode contact or a dry field, a damaged handpiece / cable, a return-electrode (REM/NESSY) fault, or an output-stage fault.
Symptoms: REM / NESSY / plate alarm, monopolar output inhibited.
Root cause: Poorly applied or dried split dispersive pad, a pad-cable fault, or the contact-quality-monitor circuit.
Symptoms: Bipolar or sealing output weak, seal cycle will not complete, instrument not recognised.
Root cause: Bipolar cable / instrument fault or the sealing output stage.
Symptoms: Output power or the REM circuit outside tolerance on the analyser.
Root cause: Output board or REM circuit needs recalibration - anti-static handling required.
Symptoms: Output will not activate from the pedal or the handpiece.
Root cause: Footswitch or handswitch / cable fault.
Symptoms: Unresponsive touchscreen, will not power on or boot, fan / thermal alarm.
Root cause: Display, power supply, control board, or a cooling-fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibration required after replacement; anti-static handling.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reusable instrument / handpiece cables | Periodic replacement | $$ | Easy |
| Split dispersive (return) pads | Consumable | $ | Easy |
| Footswitch | Periodic replacement | $$ | Fairly easy |
| Bipolar / sealing cable | Periodic replacement | $$ | Easy |
| RF output board Recalibration required after replacement; anti-static handling. | Major service part | $$$$ | Difficult |
| Cooling fan | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.