UNICO · Model IV952
Inverted Microscope, 4-Hole Nosepiece. Intensity Rheostat for Lighting Control. Plan Achromat, or Plan Phase Objectives. Left Eyetube with Diopter Focus Control. Graduated Mechanical Stage, Low Position Controls. Center Stage Ring Diameter is 110mm. Head: Binocular Head Inclined 45, Vertical Trinocular. IV952: Binocular Head. IV952T: Trinocular Head. Objectives: Brightfield Objectives: 4X and 10X Plan. Phase Objectives: 20X and 40X Plan Phase. Illumination: Bright 12 Volt / 30 Watt Halogen Bulb. Eyepiece: 10x Extra-Widefield Eyepieces, Paired. Condenser: N.A. 0.65. Diaphragm: Iris. Stage Size: 242mm x 172mm. Interpupillary Distance Range: 50-75mm. Focusing: Coaxial. Optional: Cross-Hair Reticule, 0.1mm Scale.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Vignetting, dark edges, uneven brightness across the field.
Root cause: Condenser height/aperture wrong, field diaphragm not centred, nosepiece not clicked in, or dirty optics; lamp/LED aging.
Symptoms: Blurry or low-contrast image.
Root cause: Wrong or missing 0.17 mm cover glass, immersion oil on a dry objective, air bubbles in the oil, or condenser aperture mis-set.
Symptoms: No or dim light, will not brighten.
Root cause: Lamp / LED at end of life, or a rheostat / driver fault.
Symptoms: Focus slips or binds, stage stiff or slips, nosepiece detent worn.
Root cause: Worn focus clutch or dried stage grease.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Illuminator lamp / LED module | Periodic replacement | $ | Fairly easy |
| Rheostat / illuminator driver | Major service part | $ | Moderate |
| Focus mechanism / clutch kit | Major service part | $$ | Difficult |
| Objective lens (DIN achromat) | Major service part | $$$ | Fairly easy |
| Mains fuse | Major service part | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio