UNICO · Model IP750
Head: Seidentopf Binocular Head with a 80/20 Beam Splitter. Features a 55-75mm Interpupillary Distance Adjustment. Heads rotate 360, One Eye Tube with Diopter Adjustment. Nosepiece: 4-Hole Reverse Nosepiece provides the user with easier Slide Insertion and. Manipulation. The Precision Ball Bearings and Internal Stop assure operational ease and durability. Eyepiece: High Eyepoint Widefield 10x/20mm or 15X Eyepieces. Objectives: 4X, 10X, 40XR and 100XR Infinity Objectives. IP750: Achromat Objectives. IP750PL: Plan Achromat Objectives. Focus: The Low Position Coaxial Coarse and Fine Focusing System designed to lessen user fatigue. The Coarse Focus Tension Control allows to increase or decrease the Focusing Tension. The Fine Focusing is graduated at 0.002 mm. Stage: Large Heavy Duty, Built-In Mechanical Stage with Low Coaxial Controls provides larger (180mm x 150mm) Platform with X-Y Travel of 76 mm x 50 mm. Graduated in 0.1 mm assures flexibility. Condenser: Focusable and Centerable Abbe Condenser NA 1.25 with Graduated Iris Diaphragm and Filter Holder. Illuminator: The Built-In Illuminator with Field Diaphragm provides Koehler Type. Illumination. The Brightness Intensity Control provides steady even illumination at all levels. Variable Intensity Rheostat Wheel is located on the left side of the base. Electric: UL & CSA Approved Electrical Parts, Three Wire Grounded Electric Cord. 110V-60Hz to 240V-50Hz. Shipping Size: 11in x 14in x 18in. Shipping Weight: 20 lbs.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Lamp / LED out, flickering, uneven or dim field, colour-temperature shift.
Root cause: Halogen bulb or LED module at end of life, a lamp-power-supply fault, or a bad connection.
Symptoms: Coarse / fine focus slips, drifts, binds, or the stage will not hold height.
Root cause: Worn focus-drive gears or slipping tension clutch.
Symptoms: Mechanical stage or condenser will not move smoothly or hold position.
Root cause: Worn stage rack-and-pinion or a loose condenser mount.
Symptoms: Haze, spots or debris in the field of view that rotate with an element.
Root cause: Dirty objectives, eyepieces, condenser or camera port.
Symptoms: No image, poor colour, software will not connect.
Root cause: Camera, its cable, or the imaging software.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Illumination bulb / LED module | Frequent wear item | $$ | Fairly easy |
| Lamp power supply | Major service part | $$ | Moderate |
| Focus-drive gear / clutch kit | Major service part | $$ | Difficult |
| Stage rack-and-pinion | Major service part | $$ | Difficult |
| Objective / eyepiece (as damaged) | Major service part | $$$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.