Unetixs Vascular · Model MultiLab Series 2-CP
Calibrated dual-channel pulse volume recording (C-PVR®). Dual-channel photoplethysmography (PPG), using easily applied digit clips. 8 MHz bi-directional TruDop® Doppler probe. 5 MHz bi-directional TruDop® Doppler probe. Definitive quality waveforms for original diagnosis or post-intervention follow-up. Extensive reporting capabilities. Easily custom-configured test protocols and reporting. Full-page color reports and waveforms required for Medicare reimbursement. Compatibility with DICOM III/PACS standards.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Segmental cuffs will not inflate or hold, pressure leaks off, study will not complete.
Root cause: Worn cuffs / tubing / connectors, an internal pump / valve fault, or a manifold leak.
Symptoms: Flat or noisy PPG / PVR waveform, no Doppler signal.
Root cause: Worn PPG probe or Doppler probe, or a probe-cable fault.
Symptoms: Recorded pressures implausible or biased, calibration fails.
Root cause: Transducer drift or a calibration issue.
Symptoms: Study will not save, calculate ABI, or generate a report.
Root cause: Workstation PC, software, or the interface.
Symptoms: No or faint waveform printout.
Root cause: Printer or paper.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Segmental cuffs & tubing set | Frequent wear item | $$ | Easy |
| PPG / Doppler probe & cable | Periodic replacement | $$ | Fairly easy |
| Internal pump / valve manifold | Major service part | $$ | Moderate |
| Pressure transducer | Major service part | $$ | Moderate |
| Workstation PC / printer | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The current luer fittings and adaptors allow for the possibility of the air line to be misconnected to an intravenous line, which may cause an air embolism.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.