Toshiba · Model Nemio MX
Ultrasound system
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Dropout, dead elements, image banding.
Root cause: Element delamination or cable damage from repeated use.
Symptoms: Intermittent image loss, connector-wiggle artifact.
Root cause: Repeated flexing and connector strain.
Symptoms: Frozen console, acquisition timeout.
Root cause: Workstation or storage degradation.
Symptoms: Dim image, color shift, dead pixels.
Root cause: Backlight or LCD panel aging.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Handle with care; most-replaced probe assembly.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Display panel | Major service part | $$ | Moderate |
| Ultrasound transducer / probe Handle with care; most-replaced probe assembly. | Periodic replacement | $$$$ | Moderate |
| Transducer cable assembly | Periodic replacement | $$ | Fairly easy |
| Workstation / GPU | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Multiple problems including : (1) The surface temperature of the transducer tip can rise and exceed the limit set by the international standard. (2) Incorrect display of the TI indices on the display monitor. The displayed value is greater or smaller than the true TI value, and (3) Manuals contain incorrect values for MI/TI values.
Transducer Surface Temperature : It was found during testing that when using the system in the 2DCDI + PW mode, with the Scale Graph function active, that the transducer surface temperature could exceed the 27 degrees Celsius limit as specified in the IEC safety standard.