With the integration of AIDR 3D into SUREExposure controls, radiation exposure is automatically reduced before the scan, ensuring that the lowest possible dose is employed for the specific diagnostic objective irrespective of the size or shape of the patient. AIDR 3D integrated in SUREExposure 3D Adaptive can be applied to all acquisition modes for routine clinical use and is able to remove up to 50% of image noise resulting, which corresponds to a dose reduction of up to 75% by the same Standard Deviation (SD) for noise. Key features of AIDR 3D. Full integration in scan protocols for improved workflow. Dose reduction in clinical setting by up to 75%. Full automatic iterative reconstruction process. Minimal penalty in reconstruction times. Noise reduction and improved Spatial Resolution. Superb artifact suppression Lightning Speed The new console enables significant improvements in workflow, with faster reconstruction, automated processing, and remote access capabilities. With a reconstruction speed of up to 28* images per second (16 images per second for standard configuration), all the images are available in a flash, even when Toshiba’s unique 0.5 mm slice thickness is used. Furthermore, the enhanced DICOM protocol allows fast data transfer speed up to 60 images per second. Generation of MPR images can be performed automatically in accordance with the exam protocol setting. High-Speed Reconstruction. Automated Processing. SUREXtension™ (Remote Access)* Latest Applications. Cardiac Examinations*. Colon View*. Lung Volume Analysis*. Vessel View*. Fat Index View* and more.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Ring or band artefacts, excessive noise, HU non-uniformity, failed daily air/water calibration.
Root cause: Detector-channel drift or a bad detector module, a DAS fault, or air-cal / tube-output drift.
Symptoms: Tube-arc errors, exposure aborts, kV/mA faults, spits, or noisy bearing; tube pulls high heat units.
Root cause: Anode / bearing wear or vacuum degradation - the CT tube is a wear item.
Symptoms: Rotation faults, comms errors during rotation, brush dust, drive alarms.
Root cause: Worn slip-ring brushes, a drive-belt / bearing fault, or the data-transmission link.
Symptoms: Table drifts, will not index to the correct position, height or cradle errors.
Root cause: Table drive, encoder, or brake fault.
Symptoms: Recon hangs or is slow, images will not send to PACS, console errors.
Root cause: Recon-hardware or console-PC fault, or the network / storage.
Symptoms: Over-temperature warnings, exposures inhibited after a series, coolant alarm.
Root cause: Chiller pump / fan, low coolant, or a blocked heat exchanger.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Major planned replacement; track heat-unit history.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| CT X-ray tube assembly Major planned replacement; track heat-unit history. | Periodic replacement | $$$$ | Expert-level |
| Detector module / DAS board | Major service part | $$$$ | Expert-level |
| Slip-ring brush set | Periodic replacement | $$ | Moderate |
| Gantry drive belt / bearings | Major service part | $$$ | Difficult |
| Table drive / encoder | Major service part | $$$ | Difficult |
| Chiller pump / coolant | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.