Ultra Low Dose Examinations This system incorporates a variety of functions that can significantly reduce the patient exposure dose, such as AIDR, SUREExposureTM 3D, and Boost3DTM as standard. AIDR uses an iterative algorithm to reduce image noise while maintaining details and structural edges. The patient dose can be significantly reduced without negatively affecting image quality. Ultra High Patient Throughput A new computer platform for AquilionTM RX provides a 60% faster reconstruction time, allowing rapid diagnosis and promoting more efficient clinical workflow. Ultra Fast MPR Generation Generation of MPR images can be performed automatically in accordance with the exam protocol setting. Ultra Fast Data Transfer Enhanced DICOM protocol allows fast data transfer speed up to 60 images per second. Automated data transfer to multiple destination can be set in exam protocol setting. Adaptable console options - 1 or 2 Scan and display consoles maximize workflow. Both consoles can be operated independently, enabling parallel processing, significantly improving the efficiency of image processing and diagnosis. Powerful postprocessing CT application - on your own computer SUREXtension offers remote access to Aquilion’s advanced display console functions and clinical applications from your own computer on the hospital network. This thin-client display station makes it possible to review images anywhere in the hospital without relying on a PACS system, increasing patient throughput and shortening waiting times.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Ring or band artefacts, excessive noise, HU non-uniformity, failed daily air/water calibration.
Root cause: Detector-channel drift or a bad detector module, a DAS fault, or air-cal / tube-output drift.
Symptoms: Tube-arc errors, exposure aborts, kV/mA faults, spits, or noisy bearing; tube pulls high heat units.
Root cause: Anode / bearing wear or vacuum degradation - the CT tube is a wear item.
Symptoms: Rotation faults, comms errors during rotation, brush dust, drive alarms.
Root cause: Worn slip-ring brushes, a drive-belt / bearing fault, or the data-transmission link.
Symptoms: Table drifts, will not index to the correct position, height or cradle errors.
Root cause: Table drive, encoder, or brake fault.
Symptoms: Recon hangs or is slow, images will not send to PACS, console errors.
Root cause: Recon-hardware or console-PC fault, or the network / storage.
Symptoms: Over-temperature warnings, exposures inhibited after a series, coolant alarm.
Root cause: Chiller pump / fan, low coolant, or a blocked heat exchanger.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Major planned replacement; track heat-unit history.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| CT X-ray tube assembly Major planned replacement; track heat-unit history. | Periodic replacement | $$$$ | Expert-level |
| Detector module / DAS board | Major service part | $$$$ | Expert-level |
| Slip-ring brush set | Periodic replacement | $$ | Moderate |
| Gantry drive belt / bearings | Major service part | $$$ | Difficult |
| Table drive / encoder | Major service part | $$$ | Difficult |
| Chiller pump / coolant | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.