Toshiba · Model Aquilion LB
90 cm wide bore. Enabling a breast board tilt of up to 25 degrees. Radiotherapy table top* with indexed patient positioning system (IPPS). Scan range of 175 or 194** cm. Low noise direct drive for 0.5 sec/rotation. Maximum 70 cm scan Field of View with full image quality. Up to 85 cm display Field of View reconstruction for oncology applications. 0.35 mm voxel resolution. Versatile in scan slice width: 16 x 0.5 mm, 16 x 1.0 mm or 16 x 2.0 mm. Quantum detector material with highest sensitivity for best low contrast resolution (2mm at 3 HU). Powerful X-ray tube: 7.5 MHU MEGACOOL. SUREExposure, advanced body contour dose reduction technology. Quantum Denoising Software, dramatically reducing exposure dose, while maintaining image quality. Boost3D is a separate filter algorithm developed for regions with high attenuation and can be combined with other filter as standard QDS. Varian RPM camera for 4D respiratorigated imaging (prospective or retrospective). Unique full automatic 4D multiphase reconstruction* (no xml-file import required).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Ring or band artefacts, excessive noise, HU non-uniformity, failed daily air/water calibration.
Root cause: Detector-channel drift or a bad detector module, a DAS fault, or air-cal / tube-output drift.
Symptoms: Tube-arc errors, exposure aborts, kV/mA faults, spits, or noisy bearing; tube pulls high heat units.
Root cause: Anode / bearing wear or vacuum degradation - the CT tube is a wear item.
Symptoms: Rotation faults, comms errors during rotation, brush dust, drive alarms.
Root cause: Worn slip-ring brushes, a drive-belt / bearing fault, or the data-transmission link.
Symptoms: Table drifts, will not index to the correct position, height or cradle errors.
Root cause: Table drive, encoder, or brake fault.
Symptoms: Recon hangs or is slow, images will not send to PACS, console errors.
Root cause: Recon-hardware or console-PC fault, or the network / storage.
Symptoms: Over-temperature warnings, exposures inhibited after a series, coolant alarm.
Root cause: Chiller pump / fan, low coolant, or a blocked heat exchanger.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Major planned replacement; track heat-unit history.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| CT X-ray tube assembly Major planned replacement; track heat-unit history. | Periodic replacement | $$$$ | Expert-level |
| Detector module / DAS board | Major service part | $$$$ | Expert-level |
| Slip-ring brush set | Periodic replacement | $$ | Moderate |
| Gantry drive belt / bearings | Major service part | $$$ | Difficult |
| Table drive / encoder | Major service part | $$$ | Difficult |
| Chiller pump / coolant | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.