Toshiba · Model Aplio XG
Exceptional performance, simply superb image quality Aplio XG provides outstanding clinical performance backed by a comprehensive array of unique, clinically validated technologies. Its full range of high-performance imaging functions lets users visualize minute tissue details and vascular structures with precision for a faster, more accurate diagnosis. Precision Imaging Aplio XG’s Precision Imaging technology enables users to acquire ultrasonic images of unprecedented clarity and resolution and thus lets users see more detail than with any other system. Innovations for life Aplio XG's extensive breast package provides you with a full spectrum of clinical tools from the most superb image quality, high-resolution 4D and MicroFlow contrast imaging up to unique new technologies. MicroPure is an innovative clinical tool that can help users identify microcalcifications, a potential marker for malignancy in breast tumors. Toshiba’s new ElastoQ quantitative relaxation elastography assists in quantifying palpation of breast masses by providing a robust numerical indicator for malignancy. Built more intelligently to make user's work easier Featuring Toshiba’s unique and innovative iStyle package, Aplio XG is designed to support in carrying out examinations more efficiently and with an ergonomically correct, neutral posture. From the highly movable panel to the premium quality 19'' LCD screen with articulating arm and convenient handle, every aspect of Aplio XG is designed to adjust to clinical needs and personal preferences. Productivity unlimited From imaging to quantification, from reporting to archiving, Toshiba provides a full spectrum solution that helps users manage routine and advanced clinical studies more efficiently.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
Toshiba America Medical System Inc initiated a field corrective action on Aplio 50; SSA-700A; software version 5.5r002 and later, Aplio 80; SSA-700A; software version 5.5r002 and later, and Xario; SSA-660A; software version 1.0 and later, because the product does not meet the Safety Standard requirements.