TOMY · Model LC-200/220/230
Easy-Mount Rotors - Tomy offers a wide selection of Easy-Mount Rotors for the Flexpin Bench-top centrifuges. These rotors install and remove with the turn of a knob and the selection covers most of the commonly used centrifuge labware. Flex-Spin Drive - The Flex-Spin Drive features a flexible motor shaft that equalizes minor imbalances without damaging the centrifuge or triggering the imbalance protection system. This allows excellent tolerance for minor difference in sample volumes. Superior Imbalance Protection - Tomy has developed a unique concept in imbalance protection based on the kinetic energy generated by the entire rotational system including the motor, rotor, buckets and adapters. This system, with its expanded inputs, is capable of detecting a severe imbalance within a few seconds after the centrifuge is started. Higher Speeds for Microplates - The TS-38LB rotor and B438-96 buckets spin four 96-well microplates at 4,200 rpm generating 3,100G - sufficient for microplate based ethanol extractions of DNA. High-Speed Fixed Angle Rotor - The TLA-11 Fixed Angle Rotor is a high-speed rotor that takes up to six 35 to 50 ml conical centrifuge tubes, and generates up to 9,600G. The maximum speeds will depend on the tube manufacturers' specifications. Flat Tabletop - No Chamber Gasket on the tabletop provides easy sample load. Safety Devices. Imbalance Detector. Scatter Containing Lid System. Rotor Identification System. Lid Interlock System. Lid Detector. Overspeed Detector. Motor Overheat Detection. Overcurrent Detection. Circuit Detector.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.