
Thermo Scientific · Model Shandon Cytospin 4
Safe and Reliable. Auto-locking, plastic outer lid. Autoclavable Sealed Head. Disposable sample chambers with caps. Safety alarms that protect users and specimens. Wipe-clean control panel Features. Process 12 specimens at one time. Accepts all protocols from Shandon Cytospin 1, 2, and 3. Redesigned Lid-release Mechanism allows for one-handed opening and closing of the outer lid for easy access. Excellent safety design prevents accidental fluid spillage from damaging the mechanical or electronic components and allows for easy disinfecting. Pull-out Program Card allows for up to nine programs to be logged for fast and convenient retrieval (erasable ink pen included) Shandon Cytospin Sealed Head. One Sealed Head is included with the Cytospin, but many laboratories shorten preparation time by having a second Sealed Head ready for when the first finishes processing. Autoclavable, 12-chamber Head seals completely Simple and Precise Controls. Protected Program Memory stores up to 23 routines for instant recall and protects from power loss. Program Key Pad has “easy-touch” design that allows for easier keystrokes and more intuitive programming. Speed window displays the programmed spinning speeds from 200 – 2000 rpm. Three Acceleration Control buttons allow the operator excellent flexibility and exceptional protection for fragile samples. Time window displays the programmed and remaining run time from 1– 99 minutes. Control feature has a standby power saving mode. Displays are also turned on and off by the standby key Safety Alarms. Audible alarm that signals the operator if the Shandon Cytospin Cytocentrifuge is out of balance, outside the speed tolerance, or if the lid is not properly locked. As an extra measure of safety for the user, the unit will not spin if the lid is not locked. Specimen safety alarm feature has been added to remind users at one-minute intervals to remove specimens, protect them from “air-drying” and ensure consistent results.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio