Thermo Scientific · Model Heraeus Multifuge X3
Increased Capacity and Processing Throughput - Process up to 40 x 50 ml conical tubes (40% more) in a swinging bucket rotor, or up to 14 x 50 ml (100% more) in fixed angle configuration, for significantly increased capacity over competitive models in the same bench space. Broad Range of Rotors - For exceptional versatility, the Heraeus Multifuge X3 supports the broadest range of rotors and adapters available, including Thermo Scientific Fiberlite rotors, for a minimum of 16 rotor configurations. Energy Efficiency - Achieve up to 40% energy consumption savings on industry standard protocols, such as blood separation or conical tubes processing, with the Heraeus Multifuge X3. Space-Savings Design - Realize exceptional capability in compact dimensions for maximized workspace, and with ergonomic height and curved design ensure comfortable loading and unloading of samples and cleaning of the unit. Advanced, Intuitive Programmable Controls - Detailed operation information, including display of set and actual parameters, maximize run performance and temperature accuracy for improved sample protection. Easily adjust speed, time and temperature, even during the run with our convenient keyboard for direct input. Protection for Temperature Sensitive Samples - For high-performance applications with temperature-sensitive samples, the chamber and sample control is available between -10° and +40°C. With our double temperature management system, display the air temperature in the chamber as well as the sample temperature, ensuring accurate control of the sample temperature during the run. Auto-Lock® III Rotor System - A secure locking system allows easy push-button installation and exchange of rotors for quick switch between applications and time savings. Also, easy access to the chamber enables quick cleaning, a healthier working environment and longer unit life. ClickSeal® Bucket Sealing System - THERMO SCIENTIFIC's biocontainment solution, certified by CAMR in Porton Down, UK, allows simple operation eliminating screw caps and complicated clips as well as glove-friendly, one-handed open/close capability. Motorized Lid Latch - Improve closure time and reduce wear with our highly reliable, hybrid lid latch system -combining the advantages of mechanical and electromechnical systems. This patented technology enables one finger downward motion for easiest closing and locking of centrifuge lid- even on tall benches with optional automatic lid opening. SMARTSpin® Technology - The advanced rotor management system maximizes acceleration, braking and residual load imbalance for each rotor and bucket option, optimizing safety and improving separations.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.