TEKNO-MEDICAL · Model TOM 500
Clear structured front panels. Up to 99 easy retrievable memories , 89 of them individually programmable and up to 10 preset programs Two monopolar tension controlled cutting currents ( CV "constant Voltage') up to four monopolar coagulation currents , including MICRO, MACRO, and sealing for a reliable vessel sealing Bipolar coagulation with automatic Start / Stop function Active control of HF Leakage currents to prevent undesired HF irritations
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Poor cut/coag, sparking, low power.
Root cause: Burned/eroded blade or pencil.
Symptoms: Pad alarm, no activation.
Root cause: Poor contact or expired pad.
Symptoms: Fractured tip, loose component, debris.
Root cause: Metal fatigue, improper reprocessing, or misuse.
Symptoms: Residue, bioburden, discoloration.
Root cause: Inadequate cleaning cycle or damaged tray.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
High-turnover consumable instruments.
Consumable per procedure.
Single-use; expiry critical.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Table hydraulic actuator | Periodic replacement | $$$ | Difficult |
| Instrument repair / replacement set High-turnover consumable instruments. | Periodic replacement | $$ | Fairly easy |
| Electrosurgical pencil & electrode Consumable per procedure. | Consumable | $ | Easy |
| Return electrode (patient plate) Single-use; expiry critical. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →