Techno Medica · Model GASTAT 601
Replacing consumables All models in GASTAT-600 series are equipped with a barcode reader. Through their sophisticated barcode functionality, they automatically indicate when the time has come to exchange the reagents and other consumables. They also collectively manage patient ID and quality control data. Maintenance-free electrodes The newly-designed long shell life electrodes make a great contribution to high running cost-saving performance They require no maintenance, and provide dependable, high-precision measurements. Automatic aspiration system The automatic aspiration system minimizes the risk of infection during sample loading. The inlet for samples is easily replaced via a one-touch procedure. It accepts samples in either syringe or capillary.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: pH, pCO2, pO2, electrolyte, glucose / lactate or CO-oximetry results flagged, QC fails, stability / response errors.
Root cause: An electrode / membrane at end of life, an air bubble or protein in the measuring path, or a calibration fault.
Symptoms: Sample not aspirated, clot detected, inlet / preheater alarm.
Root cause: Clot in the inlet probe / preheater, or a worn inlet gasket.
Symptoms: Reagent-empty, gas-supply or waste-full alarm, calibration fails.
Root cause: Empty or expired solution pack / cartridge, an empty / expired gas cylinder, or a full waste container.
Symptoms: Poor sample transport, calibration or rinse failures, pressure alarms.
Root cause: Worn peristaltic pump tubing, a pinch-valve fault, or a pneumatic-pressure fault.
Symptoms: tHb, sO2 or dyshaemoglobin results implausible, optics / lamp error.
Root cause: Optics contamination, a lamp fault, or the haemolyser.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sensor cassette / individual electrodes & membranes | Frequent wear item | $$$ | Fairly easy |
| Solution pack / reagent cartridge | Consumable | $$ | Easy |
| Calibration gas cylinder | Frequent wear item | $$ | Easy |
| Peristaltic pump tubing / pinch valves | Frequent wear item | $ | Moderate |
| Inlet probe / preheater gasket | Frequent wear item | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.