Tecan · Model Freedom EVO
Fully automated liquid handling and robotics for life sciences and diagnostics. The Freedom EVO series is one of the most popular and reliable systems in the market, with more than 5,000 systems installed worldwide. It offers a choice of worktable capacities, each with buildingblock modularity that brings convenient scalability, proven liquid handling, easy-to-use robotics and advanced process control within reach of every laboratory scientist.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Carriers / pucks jam, stall, or mis-route; track-error message; throughput drops.
Root cause: Worn track belts / drive, a diverter / gate fault, or a carrier-sensor fault.
Symptoms: Tubes dropped or mis-placed, gripper-position error, aliquotting mis-aligned.
Root cause: Worn gripper / robot drive or an encoder / home-sensor fault.
Symptoms: Inline centrifuge imbalance, lid-lock error, will not spin.
Root cause: Drive / bearing wear, a lid-lock fault, or an imbalance-sensor fault.
Symptoms: Caps not removed / applied, aliquot volumes off, station jams.
Root cause: Worn decapper jaws, aliquot-probe / syringe wear, or a station-drive fault.
Symptoms: Analysers or the LIS drop off the line, rules not applying, samples stuck in a queue.
Root cause: The automation middleware / server or a network link.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Track belts / drive components | Periodic replacement | $$$ | Difficult |
| Robotic gripper / drive assembly | Periodic replacement | $$$ | Difficult |
| Centrifuge-module drive / lid-lock | Periodic replacement | $$$ | Difficult |
| Decapper jaws / aliquot-probe & syringe | Periodic replacement | $$ | Moderate |
| Carrier / position sensors | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Gas springs not always replaced within the interval specified.
There is a software issue that can potentially impact sample identification and traceability under specific conditions when using EVOware Add On Software: Sample Tracking -Multi Instrument V1.0 or higher, Part Number 30019971.
Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.
WIth TouchTools 3.0 after the execution of a Rich User Prompt in Display and Wait mode, door lock errors are ignored. If an operator presses "Continue" following a Rich User Prompt and the door has been left open, the script will progress (continue without warning). This could lead to the instrument moving unexpectedly and potentially cause an injury to the operator if they are physically inside t