Stryker · Model X7000
• Single Handed Cable Insertion. • RFID Tag (Insures that only Stryker Bulbs work in the X7000). • 20% Brighter than the X6000. • Core Design. • Elliptical Bulb. • SIDNE Ready. • Multiple Safety Features. • Internal Safety Check. • Ethernet Connection. • Internal Thermal Switch. • Error & Warning Codes Display. • Single Handed Cable Insertion provides Easy Set-Up for the Busy OR. • Universal Light Cord Adapter for use with any Light Cord on the Market. • Auto Shutter for regulation of brightness with Existing Video Equipment.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No light output, or dim/flickering output.
Root cause: Xenon or LED lamp module end-of-life.
Symptoms: Overheat shutdown or thermal alarm.
Root cause: Cooling fan or vent obstruction.
Symptoms: Reduced light transmission, dim image at the scope.
Root cause: Fiber bundle damage from flex or connector wear.
Symptoms: Won't adjust brightness, stuck at one level.
Root cause: Control board or potentiometer fault.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace per lamp-hour schedule.
High-turnover; keep spares.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Xenon / LED lamp module Consumable - replace per lamp-hour schedule. | Frequent wear item | $$$ | Fairly easy |
| Cooling fan | Periodic replacement | $ | Fairly easy |
| Fiber optic light cable High-turnover; keep spares. | Periodic replacement | $$ | Easy |
| Intensity control board | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.