Stryker · Model SDC HD Digital Capture System
Sending and storing images. Flexible patient file retrieval. Simultaneous still and video documentation. Easy text input. Accurate image and video labeling. Full integration with PACS systems. Writing MPEG files. Dictating separate audio files.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pressure droops or overshoots, poor distension, pressure or flow alarm.
Root cause: Pump-head / roller wear, a pressure-sensor fault, a leak in the tubing / inflow-outflow, or a blocked cannula.
Symptoms: Cassette / tubing not recognised, will not prime, leaks at the pump head.
Root cause: Cassette-detection switch or worn pump-head seals.
Symptoms: Poor outflow, cannot clear debris, outflow alarm.
Root cause: Outflow pump / valve or a blocked suction line.
Symptoms: Blade will not oscillate / rotate, weak cut, handpiece not recognised, footswitch dead.
Root cause: Worn shaver handpiece / motor, or its cable / footswitch.
Symptoms: Unresponsive touchscreen or keys, will not power on.
Root cause: Display / control board or power supply.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pump-head rollers / seals | Periodic replacement | $$ | Moderate |
| Pressure sensor / transducer | Major service part | $$ | Moderate |
| Cassette-detection switch | Periodic replacement | $$ | Moderate |
| Shaver handpiece / motor (shaver systems) | Periodic replacement | $$$ | Moderate |
| Footswitch | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Stryker is recalling various lots of the Trithion MIS Modular Distal Capture because complaints were received that specific lots of this product may disassociate during use.
Potential for weld disassociation of the cross pin, which could lead to loose or disassociated trigger mechanism on the Distal Capture Assembly.
Users may experience issues with the functionality of the SwitchPoint Infinity 2 and SwitchPoint Element that may be exhibited as a "Page Not Found" error message, missing drop down options, or a frozen touch panel.
Device will not able to assign the correct time/date stamp for saved filed after December 31, 2009. On Jan 1, 2010, the device system date for all SDC Pro 2 units will re-start from 1 Oct 2001 and begin incrementing from that point onwards. The system date will be incorrect and users will not be able to correct this issue. Problem will not affect how pictures or videos are captured using SDC Pro
Video device is not able to assign the correct time/date stamp.
Product at times experiences intermittent signal loss due to electromagnetic interference.
Patient Identification numbers and names may be applied to images from another patient due to a software issue. This failure mode can only occur when the pre-load function of the SDC ultra is used.
Assembly problems: MIS instruments (Triathlon MIS AP Sizer Body - Left; Triathlon MIS Sizer Body - Right; Triathlon MIS Femoral Adjustment Block and Triathlon MIS Modular Capture) may not assemble properly.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.