Smith and Nephew · Model Dyonics Power II
Choice of soft tissue cutting modes adapt to surgeon preferences and procedure demand. Dual hot ports enable simultaneous activation of shaver and power handpieces. Blade family auto-recognition quickly recalls preferred settings. Footswitch can be operated in variable speed or on-off modes. Provides flexibility to activate handpiece and footswitch controls during a procedure.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pressure droops or overshoots, poor distension, pressure or flow alarm.
Root cause: Pump-head / roller wear, a pressure-sensor fault, a leak in the tubing / inflow-outflow, or a blocked cannula.
Symptoms: Cassette / tubing not recognised, will not prime, leaks at the pump head.
Root cause: Cassette-detection switch or worn pump-head seals.
Symptoms: Poor outflow, cannot clear debris, outflow alarm.
Root cause: Outflow pump / valve or a blocked suction line.
Symptoms: Blade will not oscillate / rotate, weak cut, handpiece not recognised, footswitch dead.
Root cause: Worn shaver handpiece / motor, or its cable / footswitch.
Symptoms: Unresponsive touchscreen or keys, will not power on.
Root cause: Display / control board or power supply.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pump-head rollers / seals | Periodic replacement | $$ | Moderate |
| Pressure sensor / transducer | Major service part | $$ | Moderate |
| Cassette-detection switch | Periodic replacement | $$ | Moderate |
| Shaver handpiece / motor (shaver systems) | Periodic replacement | $$$ | Moderate |
| Footswitch | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The single use devices are provided sterile and do not have an expiration date on the label.
Manufacturing issue: The distal tip of the outer sheath (return electrode) may detach and become loose in the joint during use.
Product has insufficient weld between the shaft tube and tip, could result in the tip separating from the shaft tube during use.
The sterility of the device cannot be assured
Product sterility is compromised due to breach in the sterile packaging.
External battery chargers used with the Version 29 VISTA Negative Pressure Wound Therapy pumps are failing to properly charge the pump's battery.
Integrity of the remote control bag may not be sufficient to prevent tearing or opening of the bag during use.
Sterility seal of the package tray may have gaps/voids compromising the sterility of the device.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio