Spacelabs · Model Ultraview SL2800
Seamless information exchange The Spacelabs monitoring network supports data acquisition from Ultraview patient monitors and ancillary measurement devices as well as exchange with other clinical applications on the hospital’s network. It can be scaled to support multiple care environments, an entire hospital, or even a multi-site hospital enterprise. Point-of-care access to patient information Dynamic Network Access (DNA) provides easy and immediate access to a broad range of clinical applications hosted on the hospital’s Citrix servers. Using DNA, clinicians can check lab results, enter orders, review medication guidelines and hospital protocols, perform charting, and more at the bedside monitor. DNA supports clinical decision making by making access to information convenient – all without interrupting patient care. Diverse alarm management solutions A range of surveillance features and alarm options helps to assure optimum patient safety while maximizing caregivers’ efficiency. Alarm Watch can immediately alert you to alarm events for any monitored patient at any monitor within the surveillance group. Remote View or Full Bed Review can help you assess the patient’s clinical condition remotely to determine whether immediate attention is required. Alarm Limit Review (available with specific modules) facilitates information sharing among care teams by providing a view of all bedside alarm settings. And, with the ICS Clinical Event Interface, alarms and waveforms can be instantaneously sent to personal communication devices to speed response.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: A bed shows no ECG or dropped beats at the central station, noisy trace, frequent leads-off.
Root cause: Worn transmitter lead set, a weak transmitter battery, or antenna / receiver coverage.
Symptoms: Dropout in specific areas of the unit, multiple beds affected in one zone.
Root cause: Antenna cabling, a splitter / amplifier, or a receiver-module fault.
Symptoms: Central station hangs, will not admit, alarms not annunciating, display frozen.
Root cause: Central-station PC / server, the application, or the network link.
Symptoms: No strip on alarm or on demand.
Root cause: Central recorder / network printer fault.
Symptoms: Timestamps wrong or drifting, full-disclosure gaps.
Root cause: Network time source or connectivity.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Telemetry transmitter lead set | Frequent wear item | $$ | Easy |
| Transmitter battery (per pack) | Consumable | $ | Easy |
| Receiver module / antenna components | Major service part | $$$ | Difficult |
| Central-station PC / server | Major service part | $$$$ | Difficult |
| Network printer / recorder | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.