Spacelabs · Model Ultraview SL 91387
Receiver Module.
No documents available for this device yet.
Synced live from the FDA openFDA API.
The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and discovered 222 additional serial numbers were missed and were not included in the previous recall under RES 59456 (recall number Z-3032-2011).
The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally.
The circuit that supplies backup power to the modules in the Medical Patient Monitor, Model 91387, is not working. If there is a power interruption, the three-minute backup may not be operational and customers may lose any changes they have made to the alarm settings.