Sorvall · Model RC-4
Shorter processing times for more samples - The RC-4 is designed to spin larger numbers of tubes at higher RCFs. Its powerful induction drive and swinging bucket rotor - which accommodates DoubleSpin buckets - combine to provide higher ‘g’s and capacities for the most popular sample formats. Spin more tubes at higher RCFs - The RC-4 offers exceptional capacity for all popular sample formats, including 15 and 50 ml tissue culture tubes, and many others. Spin even DNA and RNA filters plates under biocontainment at 5,890 x g! High capacity processing for all labs - With its wide range of rotors, the RC-4 is equally at home in biochemistry, biotechnology, microbiology, pharmaceutical and blood processing labs. Safe operation - The RC-4™ meets stringent international safety standards. It offers the widest selection of independently tested biocontainment rotors that comply with rigorous international guidelines for aerosol containment of bacteria and viruses. Optimum run stability - The SMARTspin™ Imbalance Detection System will automatically brake the centrifuge if an excessive imbalance is detected, for example, due to tube leakage. Reliable imbalance detection means that even large swinging-bucket rotors can operate safely with a 30 gram imbalance. Easy to use - Unlike other models, the EASYcontrol™ II Soft-Touch Keypad is located at the front of the instrument, eliminating awkward reaching. Enhanced user safety - choose from 5 biocontainment rotors accommodating samples from 1.5 - 1,000 ml. Large 4 liter/8 blood bag capacity. Fastest processing and highest throughput of any 4 liter low speed centrifuge. Unmatched ‘g’s with fixed-angle rotors for bacterial, tissue culture and general pelleting needs. Accepts 32 standard microplates or 8 deepwell plates. HIGHplate® rotor spins DNA/RNA filter plates to 85 mm in height. Optimum run stability with SMARTspin™ Imbalance Detection System. Easy-to-clean soft-touch keypad. Reduced maintenance with brushless induction drive.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.