Sorvall · Model Legend 23R
Maximum productivity. The Legend 23R excels in a wide range of applications including nucleic acid purification, PCR cleanup, plasmid mini-prep, protein processing and forensics. Its flexibility with its extensive rotor library (12 rotors) extends the capabilities of this unit. Ideal for shared labs. Auto-lock feature allows for easy rotor exchange. Ideal for users who convert between microplate and microtube rotors. Provides significant time savings that increases productivity - so you can accomplish much more in less time. High Performance Refrigeration System. Ensures samples are kept at 4C even when unit is run at maximum speed. This precise temperature control ensures no sample degradation and better quality of separation leading to improved results. Exceptionally quiet with a 57dBA, so you can conduct your research all day long without distraction. Excellent Value - The addition of the FiberLite 48 place rotor provides the Legend 23R with the competitive edge – higher capacity and performance ability, allowing you to spin more tubes at a faster rate- resulting in savings of time and money. Protection and Assurance - CAMR- approved biocontained accessories Protection and assurance for users that spin toxic or infectious materials desiring aerosol containment.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.