Siemens Healthineers · Model Rapidpoint 400/Rapidpoint 405
Rapid turnaround times and lab quality results. Easy to use, reliable, and maintenance-free operation. Simple menus guide the user through analyzer operation. Seamless integration with your LIS/HIS or with the RAPIDComm® Data Management System. Vital Stat values in just one minute. Automatic, single sample port adapts to syringe and capillary samples. Intuitive menu-prompted operation with color touchscreen panel. Cartridge-based system assures interruption free operation. All reagents contained in a single measurement and wash/waste cartridge, no gas tanks or reagent bottles. Clot detection and clearance with biosafe automatic sampling system. On-board automatic quality control cartridge requires no operator intervention during the 28-day use life. No routine maintenance. Planar sensor technology ensures industry proven accuracy and reliability. Centralized data management through seamless integration with our RAPIDComm Data Management System. POC analyzers can be monitored and controlled from a single RAPIDComm workstation. Instant clinician access to patient results from networked POC analyzers.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: pH, pCO2, pO2, electrolyte, glucose / lactate or CO-oximetry results flagged, QC fails, stability / response errors.
Root cause: An electrode / membrane at end of life, an air bubble or protein in the measuring path, or a calibration fault.
Symptoms: Sample not aspirated, clot detected, inlet / preheater alarm.
Root cause: Clot in the inlet probe / preheater, or a worn inlet gasket.
Symptoms: Reagent-empty, gas-supply or waste-full alarm, calibration fails.
Root cause: Empty or expired solution pack / cartridge, an empty / expired gas cylinder, or a full waste container.
Symptoms: Poor sample transport, calibration or rinse failures, pressure alarms.
Root cause: Worn peristaltic pump tubing, a pinch-valve fault, or a pneumatic-pressure fault.
Symptoms: tHb, sO2 or dyshaemoglobin results implausible, optics / lamp error.
Root cause: Optics contamination, a lamp fault, or the haemolyser.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sensor cassette / individual electrodes & membranes | Frequent wear item | $$$ | Fairly easy |
| Solution pack / reagent cartridge | Consumable | $$ | Easy |
| Calibration gas cylinder | Frequent wear item | $$ | Easy |
| Peristaltic pump tubing / pinch valves | Frequent wear item | $ | Moderate |
| Inlet probe / preheater gasket | Frequent wear item | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.