Shimadzu · Model SDU-1200 Pro
Micro Imaging Array Technology Provides Ultra-High-Definition Images Delivers the highest level of spatial resolution available via our unique Micro Imaging Array technology. It realizes high frequencies and wide aperture imaging at up to15 MHz to expand the boundaries of ultrasound diagnosis. Advanced Digital Technology Delivers High Image Quality The Pro is equipped with a digital beam former that uses advanced digital technology to deliver higher-quality ultrasound images. The SDU-1200Pro full-digital color diagnostic ultrasound system faithfully records a large amount of diagnostic information and provides the high frame rates, accurate focus, wide dynamic range, and high frequencies demanded for ultrasound diagnosis. True Motion Imaging Captures Transitory Movement The SDU-1200Pro's high frame rate captures even momentary movement during examinations. True Motion Image provides high temporal resolution for greater diagnostic confidence. ASI Displays an Optimized Image at the Touch of a Single Key Commonly used parameters can be stored in the ASI (Application Specific Imaging) key. Even in CFM mode that requires complex parameter settings, a single key touch displays images using previously-registered image parameters. Shimadzu's SDU-1200Pro lets anyone easily register the image parameters needed for their own clinical applications.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.