Shimadzu · Model MobileDaRt Evolution EFX
Low chassis height ensures excellent forward vision. Drive away immediately; no waiting around for the PC to boot-up before drive is possible. SSD for very fast PC / DR system boot-up time. With battery powered drive, integral independent drive motors, along with sensitive steering & speed enable effortless and accurate positioning of the MobileDaRt in confined spaces. The tube support column rotates 270o and the ergonomic telescopic arm allows smooth, fast positioning of the X-ray tube. The ‘Inch mover’ facility allows fine tuning of the unit’s position from the tubehead. The optional second exposure switch allows the radiographer to acquire an image when they are positioned remotely to the trolley or bed. Optional Keyless Entry with 10 different passwords customises individual user setup and protocols and means no more lost keys. Optional wireless remote handswitch allowing collimator and exposure activation. Back-up detector cable if wireless communication is not possible or is blocked.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Exposure inhibited, kV / mAs error, images under- or over-exposed.
Root cause: HV generator / capacitor fault, exposure-switch fault, tube fault, or a thermal interlock.
Symptoms: Unit will not drive, drives weakly, or a drive fault code.
Root cause: Drive battery, drive motor / controller, or the throttle handle.
Symptoms: Few exposures per charge, will not reach full charge, unit dies mid-round.
Root cause: Aged exposure / drive battery bank or a charger fault.
Symptoms: Tube arm drifts, will not extend / rotate, or a brake will not hold.
Root cause: Worn arm bearings, counterbalance, or brakes.
Symptoms: Panel not detected, images not received, console errors.
Root cause: Wireless detector, its battery, or the acquisition PC.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery bank (exposure / drive) | Periodic replacement | $$$ | Moderate |
| HV generator / capacitor module | Major service part | $$$$ | Expert-level |
| Drive motor / controller | Major service part | $$$ | Difficult |
| Column arm brake / bearing kit | Major service part | $$ | Moderate |
| Wireless detector / battery (DR) | Periodic replacement | $$$$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
There is a potential that the adjustable handle option on the X-Ray system may have been installed with improper bolts. This could result in improper driving operation which could lead to harm of individuals around the device.
In these units, the brake is usually released by gripping the drive handle and applied by letting go of the handle. However, for systems within a specified manufacture period, we have identified that even when letting go of the handle, the systems brakes are not applied. In this event, if the Operator releases the handle when the device is moving, the unit will continue to coast.
The firm is recalling the Shimadzu Mobile X-ray system because an image may not transfer to image server properly.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.