Samsung · Model GM85
S-Align - provides precise alignment for superior imaging SID (Source to Image Distance) Guide - supports multiple SSID settings to aid precise device positioning Multi-touch - allows users to control and adjust images easily and intuitively with fingers only Detector charging and storage - radiographers can charge detector and detector battery while traveling and Detector locking function secures detectors to prevent loss.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No exposure on the exposure switch, error / alarm message, exposure inhibited, images under- or over-exposed.
Root cause: HV generator fault, an exposure-switch / hand-switch fault, an HT-cable fault, or a system-battery fault.
Symptoms: Motorised drive weak or dead, few exposures per charge, low-battery / drive alarm.
Root cause: Aged battery pack or a drive-motor / charger fault.
Symptoms: Detector not connecting, dropped images, poor image quality, calibration overdue.
Root cause: Detector battery / Wi-Fi link, detector damage, or a gain / offset calibration issue.
Symptoms: The arm drifts or will not hold position, the collimator light is out, tube-angle readout wrong.
Root cause: Worn arm counterbalance / brake, loose fasteners, or a collimator lamp / potentiometer fault.
Symptoms: Unresponsive console keys, intermittent operation.
Root cause: Damaged membrane switch assembly or the console-to-generator cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$$ | Fairly easy |
| X-ray tube / HV generator components | Major service part | $$$$ | Expert-level |
| HT cables | Major service part | $$$ | Difficult |
| Wireless DR detector | Major service part | $$$$ | Fairly easy |
| Console membrane switch assembly | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.