Shimadzu · Model MobileDaRt Evolution
Light-touch, natural maneuvering A light pressure applied to the drive handle moves the mobile system in a natural manner as if it were an extension of one's hand, allowing the MobileDaRt Evolution to be controlled easily and effortlessly. "Inch-mover" The MobileDaRt Evolution can be moved forward or backward by using the bedside drive controls located on the front of the collimator. Safety is a major consideration and any sudden force applied to the handle during "Inch-mover" operation stops the MobileDaRt Evolution automatically. In addition, X-ray irradiation is also automatically disabled during any movement of the MobileDaRt Evolution. Keyless Entry Setting a password for each user enables keyless use of the system. The Key-less Entry enables to turn on the system by 4 digits key command input without any actual key, which is useful to share the units by several staffs. Wireless Networking Standard MobileDaRt Evolution features include support of DICPM print and storage. Output data in DICOM format to a laser printer, image server, or viewer using the provided LAN connection. Optional wireless network connection is available that provides even greater freedom for communication to RIS or PACS while moving the unit.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Exposure inhibited, kV / mAs error, images under- or over-exposed.
Root cause: HV generator / capacitor fault, exposure-switch fault, tube fault, or a thermal interlock.
Symptoms: Unit will not drive, drives weakly, or a drive fault code.
Root cause: Drive battery, drive motor / controller, or the throttle handle.
Symptoms: Few exposures per charge, will not reach full charge, unit dies mid-round.
Root cause: Aged exposure / drive battery bank or a charger fault.
Symptoms: Tube arm drifts, will not extend / rotate, or a brake will not hold.
Root cause: Worn arm bearings, counterbalance, or brakes.
Symptoms: Panel not detected, images not received, console errors.
Root cause: Wireless detector, its battery, or the acquisition PC.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery bank (exposure / drive) | Periodic replacement | $$$ | Moderate |
| HV generator / capacitor module | Major service part | $$$$ | Expert-level |
| Drive motor / controller | Major service part | $$$ | Difficult |
| Column arm brake / bearing kit | Major service part | $$ | Moderate |
| Wireless detector / battery (DR) | Periodic replacement | $$$$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
There is a potential that the adjustable handle option on the X-Ray system may have been installed with improper bolts. This could result in improper driving operation which could lead to harm of individuals around the device.
In these units, the brake is usually released by gripping the drive handle and applied by letting go of the handle. However, for systems within a specified manufacture period, we have identified that even when letting go of the handle, the systems brakes are not applied. In this event, if the Operator releases the handle when the device is moving, the unit will continue to coast.
The firm is recalling the Shimadzu Mobile X-ray system because an image may not transfer to image server properly.
The recall was initiated because Shimadzu Corporation has identified a potential issue in certain manufacturing lots of the Shimadzu MobileDaRt Evolution in combination with CXDI-70C, CXDI-55C, and CXDI-55G. As a result of Shimadzu's investigation, there is a potential risk for screws becoming loose inside the Shimadzu Mobile X-ray systems.