Shimadzu · Model MobileArt Plus MUX-100H
"Status Indicator" You can easily recognize ready time for the exposure by color of Status Indicator. It reduces the time for patient to hold the breath. "Telescopic Arm" with "All free button" Telescopic arm provides easiest and accurate positioning. And the "All free button" in the middle of arm enables a operator to extend the arm from the system side, without going around the bed. "Inch Mover" Button allows you to make very precise movement the system to adjust the position along bed side, without moving the patient at all.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No exposure on the hand switch, error code, images under- or over-exposed, exposure inhibited.
Root cause: HV generator fault, a hand-switch / exposure-relay fault, an HT-cable fault, or a battery / capacitor-charge fault.
Symptoms: Few exposures per charge, powered drive weak or dead, will not charge, low-battery alarm.
Root cause: Aged battery pack or a drive-motor / charger fault.
Symptoms: Blurry, noisy or low-contrast images despite correct technique.
Root cause: Tube focal-spot degradation, generator instability, HT-cable insulation, or the DR detector / wireless link.
Symptoms: The tube arm drifts, will not hold position, or the column brake is ineffective.
Root cause: Worn arm counterbalance / brake, or loose fasteners.
Symptoms: Collimator light out or dim.
Root cause: Collimator lamp at end of life.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Discharge capacitors and disconnect batteries before service.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$$ | Fairly easy |
| X-ray tube / HV generator components Discharge capacitors and disconnect batteries before service. | Major service part | $$$$ | Expert-level |
| HT cables | Major service part | $$$ | Difficult |
| Wireless DR detector | Major service part | $$$$ | Fairly easy |
| Column / arm counterbalance & brake kit | Major service part | $$ | Moderate |
| Collimator lamp | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Shimadzu Corporation is recalling the Shimadzu Mobile X-ray Sytems (MUX-100 and MUX-100H) because the connecting metal plate which connects the main chassis and the column of MUX-100 may crack.