Scilogex · Model SCI124R
Large LCD display provides all visual information Bio-safe aluminum rotor/anti-sample overflow prevents accidental leakage during centrifugal process allowing risk-free microbial contamination Comes with a 24 place rotor, 1.5/2.0mL tubes. Optional adapters for 0.2mL and 0.5mL tubes Pre-cooling to desired temperature ensures optimal protection of the sample before warming Guaranteed 2°C cooling within only 7 minutes of both chamber and samples even at maximum speed and capacity utilization Sound-alert allows your undivided attention on other tasks warming Quiet running, noise level ≤ 54dB Lid lock facility allows for internal temperature maintenance even after centrifugation Intermittent temperature control mode, CFC-free refrigerant and low carbon life rotor capacity Brushless motor drive, maintenance free High quality aluminum alloy bio-containment rotors, safe and autoclavable for unlimited service life Special double cooling air design, reduces rotor heating -20°C to 40°C temperature range Powerful compressor and cooling system, fast cooling Optional rotors below include a 4 x 8-tube PCR strip rotor Pulse operation for quick spins Interlock automatic door lock Over-speed detection Over-temperature detector cTUVus approved for UL and CSA.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio