
Sartorius · Model SIGMA 8K
Compact refrigerated floor standing centrifuge for universal use in blood banks, clinical laboratories and research. Capacity up to 6L in angle rotor 12 × 1L, 6 × 1,5L or 12 blood bags in swing-out rotor. Solid construction, suitable for continous use. Maintenance-free drive, high imbalance tolerance up to approx. 80 g. Speed range 100 – 10.500 rpm. Automatic rotor identification prevents spinning rotors overspee. Efficient refrigeration; temperature adjustable between -20 °C and + 40 °C, standstill cooling, with refrigerant R 404a. Stainless steel chamber. The lid is hinged on the left hand side. Easy opening of centrifuge lid via pneumatic springs. Two motorized lid locks provide safe, assured lid-locking. Emergency lid opening on power failure without opening of housing. Ergonomic loading height of just 88 cm. The user friendly control unit is integrated in the front panel. Centrifuge fits through a door opening of only 82 cm. The centrifuge stands on movable castors and adjustable feet to even out the floor. Easily serviceable. All international standards are met, e.g. IEC 1010, CE, WEEE Options. Data logging via RS 232. Water-cooled compressor. Sealed housing and watercooled compressors, sterile room version. Inert gas injection with analysis of the oxgen content in the exhaust emission.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio