Sakura · Model Tissue-Tek TEC 5
Embedding system The Tissue-Tek TEC 5 Tissue Embedding Console System consists of two modules that together create an ergonomic workstation for the production of paraffin blocks. The user can position the individual modules from either left to right or right to left, depending on personal preference. The work surfaces of both modules are at the same height and thus the specimens can be transferred easily from one side to the other. The system is fully programmable, with automatic start-up and shutdown capability and precise temperature control of all areas. The capacity of the workspace is 10 cassettes at each side of the dispensing control unit. An electric forceps can be connected easily and placed in one of the six heated wells (to prevent paraffin sticking to the forceps).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Paraffin reservoir, work surface, heated wells or cold plate off set point; blocks embed poorly.
Root cause: Failed heater element or cold-plate cooling, a temperature sensor drift, or a control fault.
Symptoms: No or weak paraffin flow, foot-switch dispense not working, drips.
Root cause: Clogged paraffin filter/strainer, a stuck dispense valve, or wrong melt temperature for the paraffin in use.
Symptoms: Blocks do not set firmly, long cool time.
Root cause: Peltier/refrigeration degradation or a fouled heatsink/fan.
Symptoms: A heated well cold, paraffin sticking to the forceps, forceps not heating.
Root cause: Failed well heater or the electric-forceps connection.
Symptoms: Console does not warm up or shut down on schedule, program lost.
Root cause: Real-time clock / control-board fault or a software issue.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Routine consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Paraffin filter / strainer Routine consumable. | Frequent wear item | $ | Easy |
| Paraffin dispense valve | Periodic replacement | $$ | Moderate |
| Reservoir / work-surface heater element | Major service part | $$ | Moderate |
| Cold-plate Peltier / cooling assembly | Major service part | $$$ | Difficult |
| Heated-well heater / electric forceps | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
This Lot has variations in melting temperature.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.