Ross · Model Flexiflo Patrol
Unique Safe-T-Valve feature on feeding set designed to prevent accidental product free flow if set is improperly loaded or becomes dislodged from pump. Free-flow alarm alerts caregiver to free-flow condition. Lock-out feature designed to prevent unauthorized changes. Special holder for safe, convenient storage of feeding set cover when not in use. Flow rates of 1 to 300 mL/hr in 1-mL increments. Dose feature 1 to 9999 mL in 1-mL increments. Memory retention for RATE, DOSE, and VOLUME FED. Unique programmable HOLD of 1 to 90 minutes. Easy-to-use dial control and display panel.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: The pump alarms and displays an error message / code and stops.
Root cause: A subsystem fault indexed in the service manual - motor / drive, a sensor, set-detection, or a board.
Symptoms: Repeated occlusion alarms, delivery stops, or an occlusion is missed.
Root cause: Kinked or clamped line, a closed roller clamp, a positional patient tube, pressure-sensor drift, or an improperly loaded set.
Symptoms: No or inaccurate feed delivery, set not detected, cassette / rotor error.
Root cause: Worn rotor / pump mechanism or a mis-loaded feeding set.
Symptoms: Delivered volume outside tolerance on the annual test, especially at low rates.
Root cause: Worn pumping mechanism / plunger drive.
Symptoms: Short run time, low-battery early after a full charge.
Root cause: Aged battery pack.
Symptoms: Set will not load, door will not latch, plunger grip poor.
Root cause: Worn door latch, plunger jaws, or anti-free-flow mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Pumping mechanism / rotor / plunger drive | Major service part | $$$ | Difficult |
| Pressure / air-detector sensor assembly | Major service part | $$ | Moderate |
| Door / latch / anti-free-flow assembly | Major service part | $$ | Moderate |
| Drive motor / gearbox | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
The enteral pump set may have a 1-3 mm slit in the silicone tubing located between the blue connector and the stretch band Safe-T-Valve.
Pumps were manufactured with an incorrect circuit board that may cause an increase in the feed rate and flush frequency of fluids.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio