Roche · Model URISYS 2400
Innovative “load-and-go” convenience. The unique, ready-to-use cassette with 400 onboard test strips Easier, faster, and more convenient loading of test strips Enhanced walkaway time of 20 minutes/up to 75 samples Once-a-month calibration.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Strip results flagged or biased, QC out of range, colour blocks misread.
Root cause: Expired or mishandled strips, a dirty / aged optics head, or a calibration issue.
Symptoms: Strip not transported, jams, or read at the wrong position.
Root cause: Worn strip-transport belt / tray or a position-sensor fault.
Symptoms: Repeated optics errors, non-uniform reads across the pads.
Root cause: Contaminated or aged reflectance LEDs / optics.
Symptoms: Poor particle images, flow-cell or camera errors.
Root cause: Flow-cell contamination, a pump fault, or the camera / focus.
Symptoms: No printout, results not sent to the LIS.
Root cause: Printer or the interface configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Strip transport belt / tray | Periodic replacement | $$ | Moderate |
| Optics / reflectance head | Major service part | $$ | Moderate |
| Calibration strips / QC | Frequent wear item | $ | Easy |
| Sediment flow cell / pump tubing (if fitted) | Frequent wear item | $$ | Moderate |
| Printer / paper | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Limits of Detection (LoD) for protein, nitrite, leukocytes, and erythrocytes on the Urisys 1100 urine analyzer with Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA test strips were higher than what is listed in the test strip method sheets may lead to false negative results for the affected parameters