Puritan Bennett · Model Puritan Bennett MA1
The Bennett MA-1 ventilator is a volume-cycled, constant flow generator that can act as an assistor, controller, or assist-controller. It is one of the most commonly used ventilators in clinical practice. With this unit, inspiration continues until a preset tidal volume is delivered to the patient--unless impedances to gas flow increase system pressures to a preselected limit. Thus, setting the maximum inspiratory pressure too low limits the ability of the ventilator to deliver the tidal volume, causing it to function as a pressure-cycled device. Other basic controls allow you to establish the sensitivity of the ventilator to spontaneous breathing attempts, the maximum flow rate, the frequency of respirations, and the oxygen percentage. Special controls permit delivery of a sigh breath and slowing of exhalation.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Unable to reach set pressure, low-flow alarm, loud noise.
Root cause: Worn bearings or failed blower motor.
Symptoms: Unexpected shutdown, persistent low-battery alarm.
Root cause: Aged internal battery or failed PSU.
Symptoms: Leak alarm, low circuit pressure, volume loss.
Root cause: Disconnected or punctured patient circuit / humidifier.
Symptoms: FiO2 out of range, oxygen alarm.
Root cause: Failed solenoid or proportional valve in blender.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Most-replaced wearable part; stock spares.
Consumable, replace per PM schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| O2 sensor (galvanic/paramagnetic) | Frequent wear item | $$ | Fairly easy |
| Exhalation valve assembly Most-replaced wearable part; stock spares. | Frequent wear item | $$ | Fairly easy |
| Flow / differential pressure sensor | Periodic replacement | $$$ | Moderate |
| Bacterial / HEPA filter Consumable, replace per PM schedule. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Shaft breaks so that adequate patient specimen sample cannot be obtained.
Directions for Use-Puritan Bennett has determined that a service interval should be defined for the device. The firm has also determined that an exhalation filter should be employed when the device is in use.
Manufacturing defect at contract manufacturer, employee error, poor training, improper inspection.