Pronk Technologies · Model SimCube Advanced
NIBP Simulations. Adult (120/80 mmHg), Neonatal (70/40 mmHg), Hypertensive (190/120 mmHg), and Hypotensive (80/40 mmHg) NIBP Simulations. Digital Manometer. Over Pressure (Peak Detect) Test mode measures the peak pressure produced by the patient monitor. The readout latches the highest pressure valve sensed, in 0.1 mmHg steps. Leak Rate Test Mode ECG/Respiration Simulations. 12-lead ECG Simulation. 8 Selectable Heart Rates: 30, 60, 90, 120, 140, 160, 200, 240 bpm. 10 Selectable Arrhythmias: NSR w/ PVCs, VTACH, VFIB, Asys, Bigeminy, Pacer, Trigeminy, ST+, 2nd Deg. Block, Square Wave. 8 Selectable Respiration Rates: 15, 30, 45, 60, 70, 80, 100, 120 bpm. Banana Plug Interface SpO2 Simulations. Saturations in 1% increments (10-100%). Pulse rates in 1 BPM increments (25-240). Create custom preset simulations for faster Alarm Testing. Perfusion Index levels in 1% increments (10-100%). Factory Preset Simulations. [i]Masimo rainbow SET®[/i] Compatible. Automatic detection of SpO2 Manufacturer. Color Display OLED Technology IBP Simulations. 4 IBP Channels. 8 Dynamic: 120/80, 60/30, 30/10, 20/10, 125/15, 60/5, 8/2, 4/1 mmHg. 7 Static: 0, 25, 50, 100, 150, 200, 250 mmHg Cardiac Output Simulations. 3, 5 and 7 l/m. No adapter cable required Temperature Simulations. YSI 400 and 700. 25, 37, and 40 degrees Celsius.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Readings disagree with a reference standard, verification check outside tolerance.
Root cause: Normal drift - the instrument needs its scheduled factory / accredited calibration (at least every 12 months).
Symptoms: Dead unit, short battery life, will not charge.
Root cause: Depleted or aged battery, or a power / charger fault.
Symptoms: Open or intermittent readings, connection errors, a specific function fails.
Root cause: Worn or broken test leads, mains adapters, patient-lead adapters, or a sensor cable.
Symptoms: One parameter (leakage, insulation, ECG / SpO2 / IBP simulation, defib / pacer energy, ventilator flow / pressure, light / dose) errors or is out of range while others work.
Root cause: An internal source / measurement circuit fault on that channel.
Symptoms: The automation / logging software will not connect.
Root cause: USB cable / port or driver.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
At least every 12 months per the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Test-lead / adapter set | Periodic replacement | $$ | Easy |
| Battery | Periodic replacement | $$ | Easy |
| Calibration (factory / accredited) At least every 12 months per the manufacturer. | Frequent wear item | $$ | Easy |
| Communication cable | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.