Calibration Error Calculator
Compare an instrument's readings with a reference at several points, showing the error as a percentage of reading and of full scale and whether each is within tolerance.
Formula
- the true value from a traceable standard
- what the instrument under test shows
- the instrument's full-scale span
How it works
A calibration check compares the instrument with a reference at several points across its range. The error at each point is the reading minus the reference, and it is expressed against the reading or the full-scale span, which are different things: an error that is small relative to full scale can be large relative to a small reading.
Manufacturers state tolerances in one of those ways, and the choice matters at the bottom of the range. The table shows each point with its limit and whether it passes; the largest error and the count within tolerance are summarised above it.
Worked example
Five points from 0 to 200, with a tolerance of 0.5% of full scale (full scale 200).
- Limit = 0.5% × 200 = 1.0 at every point.
- The errors are +0.1, +0.4, +0.6, −0.8 and −1.1.
Four of five points pass. The reading at 200 is out by −1.1 (−0.55% of full scale) and fails.
These are the values the calculator opens with, so you can check its output against this example.
Assumptions
- The reference is much more accurate than the instrument, with a test uncertainty ratio of at least 4 : 1 and traceable to a national standard.
- Conditions are as the specification requires.
Common mistakes
- Testing only at one point.
- Applying a percentage-of-reading limit at zero, where it gives a limit of zero.
- Using a reference whose own uncertainty is not small against the tolerance.