Affiniti 70 combines outstanding performance and efficient workflow. The system's PureWave transducers are designed to increase penetration in technically difficult patients, so that one transducer can provide diagnostic confidence on easy patients as well as difficult patients. Precision beamforming delivers superb spatial and contrast resolution, outstanding tissue uniformity, few artifacts, and reduced image clutter. Tissue Specific Presets (TSPs) automatically adjust over 7500 parameters to optimize the transducer for the specific exam type, producing excellent image quality with little or no need for image adjustment. The Affiniti 70 is so intuitive that users with ultrasound experience require minimal training on system use to complete an exam. The elegantly simple control panel eliminates unnecessary buttons while keeping the most used ones nearby, and the large monitor and tablet-like touchscreen add to the system's ergonomics. At just 83.5 kg, Affiniti 70 can easily be pushed down hallways and maneuvered in tight spaces. Affiniti 70 boasts a low total cost of ownership, making it a smart investment. The modular design enhances reliability and aids rapid repair, and it consumes nearly 60% less power than its predecessor. Designed to sets users ahead and helps keep them ahead, the Affiniti 70 ultrasound system delivers innovation that responds to the needs of a busy ultrasound department. Engineered for efficiency and reliability and powered by Philips superb performance, it brings outstanding image quality, walk-up usability, and broad applicability. Affiniti 70 ultrasound system’s precision beamforming, PureWave technology, Tissue Specific Presets, and efficiency tools deliver both performance and workflow for confident throughput. A complete cardiovascular solution, it provides both adult and pediatric echocardiography (including TEE), stress echocardiography, left ventricular opacification, vascular imaging, and quantification tools. Affiniti 70 ultrasound system delivers outstanding image quality and advanced tools for all gestational ages and complex gynecological cases – in a system that is out-of-the-box usable. Its simplified yet intuitive user interface and easy access to critical features produce superb diagnostic information quickly – all in an ergonomic design, to let users work with less reach and fewer steps. Philips leverages the experiences of its customers to design Affiniti 70 to address the challenges of daily scanning. With Philips Affiniti 70, workflow meets wow. The system addresses the everyday need to scan quickly and deliver results efficiently, while incorporating those innovations that make Philips ultrasound the choice of those who demand quality images and proven clinical applications.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.