Perry Baromedical · Model Sigma 40
A 40.5” internal diameter cylinder Cool and quiet treatment compartment Patient controlled temperature Adjustable stretcher and certified for patients up to 700 lbs. Extra wide stretcher with patient safety handrails Gurney adjustments of 15°, 25°, 35°,45°, 55°, and 57 ½° Optional 22” flat screen TV with CD and DVD features Intrinsically safe two-way communications system.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Slow or no rise to treatment pressure, long equalization.
Root cause: Low supply pressure, an inlet / compression valve fault, or a control fault.
Symptoms: Chamber O2 below target at the expected time, flush / purge not performing.
Root cause: Flow / purge-rate fault, O2 sensor drift, or a supply issue.
Symptoms: Controller alerts an abnormal condition and forces a safe state.
Root cause: An electronic or pneumatic subsystem fault detected by the chamber monitoring.
Symptoms: Fine cracks in the acrylic cylinder, pressure will not hold, hiss at the hatch.
Root cause: Acrylic aging / crazing (retirement item) or a worn door O-ring seal.
Symptoms: Loud airflow noise during the plateau, patient discomfort.
Root cause: Muffler / silencer degraded or a valve issue.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Inspect for crazing and retire per the manufacturer's schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Acrylic cylinder (hull) Inspect for crazing and retire per the manufacturer's schedule. | Major service part | $$$$ | Expert-level |
| Door / hatch O-ring seal | Periodic replacement | $$ | Moderate |
| O2 sensor | Frequent wear item | $$ | Fairly easy |
| Inlet / exhaust / compression valves | Major service part | $$$ | Difficult |
| Controller / monitor board | Major service part | $$$$ | Difficult |
| Muffler / silencer | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.