• High resolution - The high resolution HD CCD video chip delivers an excellent image in HDTV 1080i quality. It is unsurpassed in detail resolution and colour reproduction, providing you with images that are true to life. • Extremely lightweight body - Olympus has produced the lightest L-shaped HD URO camera head in the world. This reduces the operator fatigue during HD minimal invasive surgery such as cystoscopy, hysteroscopy and resection, even when the procedure is lengthy. • Separated remote switches - The separated remote switches can be uniquely configured for up to three functions, e.g. freeze, enhance, zoom or NBI. Because the switches are remote from the camera, the risk of camera shake when you operate a function is greatly reduced. • Narrow Band Imaging (NBI) - NBI is a function of the new URO HD camera head and is unique to Olympus. NBI improves visibility of capillaries and of the edge of mucosal structure, which can improve diagnosis and therapy in urology and gynaecology.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No image, color shift, or visible artifacts.
Root cause: Image sensor or its driver board aging.
Symptoms: Intermittent signal or no image at the tower.
Root cause: Cable flex fatigue or connector pin wear.
Symptoms: Blurry, dark, or off-center image.
Root cause: Coupler or lens element misalignment/damage.
Symptoms: Color inaccuracy that won't correct on white-balance.
Root cause: Calibration drift or button/sensor fault.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
High-turnover; keep spares.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Camera head / sensor assembly | Periodic replacement | $$$$ | Difficult |
| Camera cable High-turnover; keep spares. | Periodic replacement | $$ | Easy |
| Coupler / lens assembly | Major service part | $$$ | Moderate |
| Camera control unit (CCU) | Major service part | $$$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.