Olympus · Model HysteroFlow
Olympus HysteroFlow Pump The Olympus HysteroFlow System provides accurate and easy measurement of fluid balance and intrauterine pressure in a closed system during hysteroscopy, thereby ensuring accurate fluid loss measurements throughout the procedure, ensuring safety and comfort.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Inaccurate flow, pressure alarm.
Root cause: Worn pump rollers or tubing raceway.
Symptoms: Reported joint/uterine pressure inaccurate.
Root cause: Pressure transducer calibration drift.
Symptoms: Fluid deficit alarm, uneven distension.
Root cause: Worn solenoid valve or occluded line.
Symptoms: Leak, or won't prime.
Root cause: Worn cassette seat or cracked fitting.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Single-use consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Disposable tubing / cassette set Single-use consumable. | Consumable | $ | Easy |
| Peristaltic pump head | Periodic replacement | $$$ | Moderate |
| Pressure / flow sensor | Periodic replacement | $$$ | Moderate |
| Inflow / outflow solenoid valve | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.