NeoMedix · Model Acquidata Median
Automated UroReport patient report with P-Q plot and nomograms. User reconfigurable display including split screen (for simultaneous comparison of flow patterns). Export of report and all waveform data to spreadsheets or databases. Simple fast two step re-calibration of transducers. All raw data saved with each file can be retrospectively reviewed using any desired settings or off-line calculations. Supplied complete with all transducers and catheters ready to start testing (user to supply bladder filling media). Remote control to allow single key operation of most functions. Phase alignment of flow trace (to perfect P-Q plots).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pves / Pabd / Pdet baselines drift, will not zero, subtracted Pdet implausible, calibration fails.
Root cause: Pressure-transducer fault, a leaking / air-bubbled line, or a transducer-dome / stopcock fault.
Symptoms: Flow trace flat, noisy, or implausible, will not calibrate to a known volume.
Root cause: Worn weight-transducer / rotating-disc flowmeter, or debris.
Symptoms: Fill rate inaccurate, pump will not run, infused-volume error.
Root cause: Peristaltic-pump wear or a control fault.
Symptoms: EMG channel noisy or dead, catheter transducer not recognised.
Root cause: EMG lead set, or the amplifier / input board.
Symptoms: Trace drops, amplifier not detected, study will not save or export.
Root cause: USB / link, driver, or the PC / software.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pressure transducers / domes & tubing | Frequent wear item | $$ | Fairly easy |
| Uroflow transducer | Periodic replacement | $$$ | Moderate |
| Peristaltic infusion pump / tubing | Periodic replacement | $$ | Moderate |
| EMG lead set | Frequent wear item | $ | Easy |
| Amplifier / input board | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.