Mortara · Model T12S
Go Wireless [b]Freedom of Movement[/b] Compact and lightweight, the T12S transmitter with SpO2 sensor provides wireless freedom of movement for the patient and clinician during ambulatory monitoring. [b]Continuous SpO2 Monitoring[/b] The T12S integrates pulse oximetry technology allowing SpO2 spot checks at the patient while continuously transmitting SpO2%, pulse rate, and plethysmographic waveform together with 12-lead ECG. The need for an external pulse oximeter device is eliminated. [b]Waveform Display and Digital Menu Navigation[/b] Full graphic display to preview ECG waveforms, check lead impedance, perform SpO2 signal quality checks during patient hookup with spot checks during transmission, channel selection, and other setup features. [b]Comfortable ECG Acquisition for the Patient[/b] Both the T12S and the LeadForm ECG cable with prepositioned leads are designed to eliminate the cumbersome inconvenience of heavy cabling and maximize comfort. [b]Diagnostic Quality[/b] Mortara wireless technology features full diagnostic ECG quality and eliminates artifact due to traditional cable movement, resulting in cleaner, higher quality tracings. [b]User-Selectable Channels[/b] The T12S offers a wide range of channels for interference-free transmission.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Baseline wander, muscle or AC noise, "lead fail", poor 12-lead quality.
Root cause: Poor skin prep, dried or lifting electrodes, or a worn patient cable / lead set.
Symptoms: Acquisition hangs, no interpretation, or a self-test failure.
Root cause: Acquisition-module fault or a software issue.
Symptoms: No or faint report, paper jam, misregistered grid.
Root cause: Thermal print head, paper handling, or (older units) the stylus.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack.
Symptoms: ECGs will not transmit to the ECG-management system / EMR.
Root cause: Network / Wi-Fi / modem configuration or the communication module.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable / 10-lead set | Frequent wear item | $$ | Easy |
| Acquisition module | Major service part | $$$ | Moderate |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Communication / modem module | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio