Full-featured and durable electrocardiograph with a range of connectivity options [b]Portable Solution[/b] Compact and lightweight, the ELI™ 250c electrocardiograph provides comprehensive functionality, with 8.5 x 11" or A4 paper, in an easy-to-use, portable device. [b]High-Resolution Color LCD[/b] High-resolution color display provides real-time preview of 12-lead ECG and post-acquisition review of acquired ECG. [b]Distinguished Mortara VERITAS™ Resting ECG Interpretation Algorithm[/b] Widely recognized Mortara resting ECG interpretation algorithm uses gender specific and adult and pediatric criteria to provide a silent second opinion for resting ECG interpretation. [b]Choice of Wireless or Traditional ECG Acquisition[/b] ELI 250c offers a choice of either the innovative WAM™ wireless acquisition module or the AM12™ acquisition module. Both include replaceable lead wires, lead fail indicator, and remote control with buttons for ECG acquisition and rhythm printing. [b]Best 10[/b] ELI 250c automatically selects and displays the Best 10 seconds of ECG signal from the five minute segment of internal full disclosure, reducing clinical review time and eliminating the need for repeat ECGs. [b]Full Keyboard[/b] Alphanumeric elastomer keyboard features dedicated "one-touch" buttons for ECG acquisition, rhythm printing, and ECG transmission/order retrieval. [b] Information Exchange[/b] Experience the benefits of bidirectional communication via internal modem, LAN, or wireless LAN with Mortara's E-Scribe™ and Athena products, as well as third party EMR, HIS, and PACS systems via XML, PDF, UNIPRO32, and DICOM32®.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Baseline wander, muscle or AC noise, "lead fail", poor 12-lead quality.
Root cause: Poor skin prep, dried or lifting electrodes, or a worn patient cable / lead set.
Symptoms: Acquisition hangs, no interpretation, or a self-test failure.
Root cause: Acquisition-module fault or a software issue.
Symptoms: No or faint report, paper jam, misregistered grid.
Root cause: Thermal print head, paper handling, or (older units) the stylus.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack.
Symptoms: ECGs will not transmit to the ECG-management system / EMR.
Root cause: Network / Wi-Fi / modem configuration or the communication module.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable / 10-lead set | Frequent wear item | $$ | Easy |
| Acquisition module | Major service part | $$$ | Moderate |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Communication / modem module | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.