Mortara · Model ELI 230
Delivers impressive 12-lead performance, while requiring minimal space [b]Value without Compromise[/b] The ELI™ 230 electrocardiograph provides complete functionality in an ultra-portable, compact unit. Easy to use, the ELI 230 is a value electrocardiograph without compromise on features such as signal quality, accurate and reliable ECG interpretation, full-size printouts and much more. [b]High-Resolution LCD Color Display[/b] Preview up to 12 leads simultaneously before printing. Heart rate, print speed and gain, filter, time, and menu selections are all visible at a glance. [b] Mortara VERITAS™ Resting ECG Interpretation Algorithm[/b] Internationally recognized Mortara resting ECG interpretation algorithm with adult and pediatric criteria provides a silent second opinion for 12-lead ECG interpretation. [b]Choice of traditional (AM12™) or wireless ECG acquisition (WAM™)[/b] The ELI 230 electrocardiograph offers a choice of either the AM12 acquisition module or the WAM wireless acquisition module, both with replaceable lead wires and remote control of the ELI 230 with buttons for ECG acquisition and rhythm printing. [b]Diagnostic Quality[/b] Mortara technology features full diagnostic ECG quality with a high-fidelity acquisition sampling rate of 40,000 samples per second per channel, a benefit for accurate pacemaker detection. [b]Efficient Power Management[/b] The ELI 230 is powered by AC or an internal battery that only needs a recharge after 200 ECGs.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Baseline wander, muscle or AC noise, "lead fail", poor 12-lead quality.
Root cause: Poor skin prep, dried or lifting electrodes, or a worn patient cable / lead set.
Symptoms: Acquisition hangs, no interpretation, or a self-test failure.
Root cause: Acquisition-module fault or a software issue.
Symptoms: No or faint report, paper jam, misregistered grid.
Root cause: Thermal print head, paper handling, or (older units) the stylus.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack.
Symptoms: ECGs will not transmit to the ECG-management system / EMR.
Root cause: Network / Wi-Fi / modem configuration or the communication module.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable / 10-lead set | Frequent wear item | $$ | Easy |
| Acquisition module | Major service part | $$$ | Moderate |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Communication / modem module | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio