Mindray · Model M7
High Speed User Experience Compact, lightweight design enables convenient access to all point of care environments 15" high resolution TFT LCD display with 170° viewing angle Fast power up Superior computing power: instantaneous response to user command and short cycle time Sealed surface for easy infection control iTouch: one button image optimization User programmable exam presets: quick start and consistency iStation: on board workstation for patient information management and connectivity Wired and iRoam, wireless, data transfer and connectivity Premium Quality Performance Through Advanced Technologies System On Chip (SOC): enables high end system technologies in an extremely compact design Octal-beam imaging technology: provides excellent temporal resolution iClear: speckle suppression technology iBeam: spatial compounding iZoom: enables accurate viewing of image for user from distance iScape: panoramic imaging TDI: tissue Doppler imaging IMT: auto-measurement of carotid intimae-thickness Full-featured DICOM function: network, worklist, MPPS, query/retrieve, structure reports Upgradeable platform guarantees long-term investment protection Flexibility to meet exchanging needs Excellent system performance across a wide range of applications User selectable setup, from dedicated specialty to full service System ergonomic and utility: hand-carried to full service cart with external display Wired and wireless data transfer and connectivity Optional system cart: height adjustable, supports expanded functionalities, peripherals and storage Expansion of current system capability Addition of new capabilities in future releases.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.