Micromed · Model MYOQUICK
• New 5, 9, 13 and 17 Channel EMG/EP system • Excellence in Signal with matrix Amplifier: - Sampling Rate up to 65536 Hz for Every Channel - 16 bit Analog/Digital Conversion • Best Quality in EMG: - Unlimited needle EMG recording and playback - Saving time customizable protocols for every exam - Automatic online report generation - Dedicated control keypad and footpedal - Maximum flexibility in Single Fiber EMG • Extreme Performance with energy Stimulator: - Up to 2 fully programmable stimulators - Up to 300 Hz electrical stimulation rate - Up to 4 stimulator outputs - Automatic stimulation artifact removal • Perfect Accuracy in Evoked Potentials: - Stimulation artifact free recordings - Fully adjustable rejection thresholds - High quality acoustic and visual stimulations - Advanced filters for better signal quality • Recombination of each channel to any input connectors on the headbox • Optional Intra Operative Monitoring (IOM) package • Network Integration with: - Centralized Data Storage - SQL Database - Interface to Hospital Database • Compatible with Magnetic and Laser Stimulators, and Full Management of Trigger IN/OUT
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Noisy / flat channels, impedance alarm.
Root cause: Dry gel, poor contact.
Symptoms: Dead channel, artifact.
Root cause: Front-end amplifier failure.
Symptoms: No stimulus delivered.
Root cause: Stimulator circuit wear.
Symptoms: 50/60 Hz interference.
Root cause: Shield/ground break.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Frequent R&R.
Consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Stimulator cable | Periodic replacement | $$ | Fairly easy |
| EEG electrode cap / cup Frequent R&R. | Periodic replacement | $$ | Easy |
| Patient electrode gel Consumable. | Consumable | $ | Easy |
| Amplifier channel module | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.