Medtronic · Model ACT Plus
Offers the reliability required in. Cardiovascular and vascular surgery. ECMO. Cardiac catheterization laboratories. Critical care units. Hemodialysis units.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ACT / APTT results shift, controls out of range, poor duplicate agreement.
Root cause: Test-chamber temperature out of the 37 C window, a plunger / flag-detection fault, or reagent-cartridge issues.
Symptoms: "Temperature" alert, tests will not start.
Root cause: Test-well heater or temperature-sensor fault.
Symptoms: Error message and priority alert on the display.
Root cause: A subsystem fault indexed in the Error Messages section of the operator's manual.
Symptoms: Cartridge or sample barcodes will not scan.
Root cause: Scanner needs its defaults set, or a scanner / cable fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Test-well heater / temperature sensor | Major service part | $$ | Moderate |
| Plunger / flag-detection assembly | Major service part | $$$ | Difficult |
| Disposable test cartridges / QC | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio