ERBA · Model ECL 760
7 measuring channels 8 incubation positions at 37°C +/- 0.2°C for reaction cuvettes 23 refrigerated and stir able positions for reagents, diluents and cleaner 27 positions for samples, primary tubes, vacuum system tubes and cups 72 positions for reaction cuvettes Pipetting needle with automatic preheating, collision protection, liquid level detection of and control of reagent volumes Volume of reaction liquid: 150-300 µl, pipetting volume: 4-200 µl Parameters: PT, APTT, TT, Fibrinogen, Coag. factors, D-Dimer, ATIII, Protein C, Protein S, Lupus Anticoagulans Built-in thermal printer RFID for security of use Quality Control: 12 QC user defined QC material Calibration: Automatic, manual, 6 point calibrations Storage capacity of more than 10 000 data curves program with automatic Levey-Jennings graph monitors system performance Comprehensive Erba Haemostasis reagent line Simple, desktop analyser Bi-directional LIS communication Barcode reader for sample identification 60 PT tests / hour Colour touch screen and the possibility to connect an external keyboard, a mouse and a printer.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: PT/APTT/fibrinogen results shift, controls out of range, poor duplicate agreement.
Root cause: Optical clot-detection lamp/detector aging, dirty cuvette optics, or reagent/temperature issues.
Symptoms: Short aspiration, no dispense, probe-position or liquid-level errors.
Root cause: Clogged or bent probe, or a clot in the sample.
Symptoms: Temperature alarm, temperature-sensitive results drift.
Root cause: Incubator heater or temperature-sensor fault.
Symptoms: High carryover, wash-station errors, poor precision.
Root cause: Wash pump, tubing, or a blocked wash station.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Clot-detection lamp / photodetector | Periodic replacement | $$ | Moderate |
| Sample / reagent probe | Periodic replacement | $$ | Moderate |
| Reaction cuvettes | Consumable | $ | Easy |
| Incubator heater / temperature sensor | Major service part | $$ | Moderate |
| Wash pump / tubing | Frequent wear item | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.